Product Overview

The MC110A sensor provides small size and a strong wideband rounded response from 30 kHz to 1 MHz.

Frequency Range:
 
2 kHz - 1 MHz
Shape:
 
B1 Box, small square
Size:
 
1” x 1” x 0.8” (25x25x20mm)
Connector:
 
BNC(f) on box
Weight (approx):
 
1 oz (30g)


About the MC series probes- 
These calibrated magnetic field sensors or probes are cost-effective tools to detect magnetic fields from 5 Hz to 1 MHz (any waveforms), and to accurately measure magnetic field strength of continuous-wave (CW sine wave) fields from 5 Hz to 1 MHz. 
These single-axis "B-dot" coil sensors respond to magnetic fields parallel to the coil axis; they produce an output voltage calibrated to the CW magnetic field strength. You display the output voltage on your own instrument (AC or RF voltmeter, multimeter, oscilloscope, or spectrum analyzer with high input impedance). The frequency of the sensor output voltage is the same as the frequency of the magnetic field. 
NIST traceable calibration certificate with data is printed and shipped with each sensor. No battery or power supply is needed for the sensors. These are easy-to-use affordable sensors for magnetic field measurement, EMC / EMI / RFI testing, and troubleshooting. These sensors do not measure static or DC magnetic fields, for example from motionless magnets or the earth's magnetic field, unless the sensor is vibrating or rotating. 
Frequency Response Graph (see photos): shows each sensor model output voltage (as Volts output per Gauss, or mV output per mG field) at each frequency, when placed in a continuous sine wave (CW) magnetic field at that frequency. Use this graph (or printed calibration data), and the sensor output voltage, to determine the magnetic field in Gauss. The graph is accurate for a continuous sine wave (CW) magnetic field. Individual sensors may vary, so calibration data is printed and shipped with each sensor.
Examples: at your frequency if the Volts per Gauss = 1 (on vertical axis), and you are measuring 80 mV, then the CW field = 80 mG.  (80mV /1 = 80mG)
           Or, at your frequency if the Volts per Gauss = 8 (on vertical axis), and you are measuring 80 mv, then the CW field =10 mG.  (80mV /8  = 10mG)
           Or, at your frequency if the Volts per Gauss = 5 (on vertical axis), and you are measuring 3 V, then the CW field =0.6 G.  (3V /5 = 0.6 G)
If your display instrument is measuring RMS volts then you are measuring RMS magnetic field strength. (RMS is recommended). 
If your display instrument is measuring Peak volts then you are measuring Peak magnetic field strength. 
Unit conversion:  10G = 1mT,  1 mG = 0.1 uT.      1 kA/m  = 1.26 mT  = 12.6 G in air.
To Use the Sensor: Connect the sensor to your display instrument (multimeter, AC or RF voltmeter, spectrum analyzer, or oscilloscope, etc). Place the center of the sensor at the location you want to measure magnetic field strength.
The sensor is single axis and responds to the magnetic field parallel to the sensor axis, which is along the longest dimension of the sensor (parallel to the writing on the sensor label). The sensing area is along most of the length of sensor axis, so if the field is spatially non-uniform it will be averaged over the length of the sensor. 
To find the maximum field polarization, turn the sensor in different directions to see the largest reading, then the sensor axis is parallel to the maximum magnetic field polarization direction. The polarization direction of the field (max reading) is often at right angles to the direction towards the source of the field. 
The reading will also increase as you get closer to the source of the field, although the immediate surroundings, multiple sources, and the sensor orientation (pointing direction) can cause variations. Sometimes you won't see exactly the same reading when you check the same location again, this is usually because the sensor is not exactly at the same location and orientation. Hold the sensor still, because vibrating, shaking, or quick movements can cause extraneous readings due to sensor acceleration through the earth’s static magnetic field.
To measure the magnetic field:  Measure the alternating (AC or RF) voltage output from your sensor, then reference the graph above (or the calibration data delivered with each sensor) at your frequency to determine the magnetic field strength. 
Minimum Measurable Field and Resolution: These are determined by the resolution and noise level of your display instrument. The sensors are passive devices and contribute negligible thermal noise.
Maximum Measurable Field: Sensors may be damaged by magnetic fields producing more than 50 Volts output from the sensor. Also, saturation of the core can cause inaccuracies above 50 Gauss (5 mT) field. The longest sensors (sizes C4 and C5) may start to saturate at 25 Gauss (2.5 mT).  Since these are B-dot sensors, a fast change in the field can cause output voltage to peak which could damage the large sensors MC90R and MC90-110 with high outputs. If unsure then best to gradually ramp-up and down the field strength. 
Temperature Range: Sensors can operate from -30 C to + 55 C (-22 F to +131 F), or in some cases a wider temp range. 
Calibration: Calibration standards and instruments are NIST traceable. Each sensor is individually calibrated using a CW sine wave magnetic field at a number of frequencies, and the calibration data is printed and shipped with each sensor. Calibration is usually accurate for many years for these passive sensors. Calibration results may vary near resonant frequencies and above 500 kHz.  
Instrument Impedance and Coax: To display the calibrated output voltage from the sensor we recommend you use an instrument with high input impedance. The length of coaxial cable you use can also significantly affect readings above 30 kHz, due to coax capacitance, see Impedance  
Sinusoidal and Non-Sinusoidal Fields: These coils are known as "B-dot" sensors since the voltage output is the time-derivative of the magnetic field. So if the magnetic field is a continuous CW sine wave, then the sensor output is a cosine CW voltage at the same frequency, and the response curves in the figure above show the correct output voltage per gauss at each frequency. 
For non-sinusoidal waveforms the output voltage is also the first derivative (B-dot) of the magnetic field. So when exposed to a rapid large change in the magnetic field (like switching the field on or off, or square wave or spike in the field), then the sensor output voltage will show a peak or bump, due to the fast change in the field giving a large first derivative. 
Some sensors are described as "resonant". Their frequency peak is the resonant frequency. Those "resonant" models will "ring" for a few milliseconds if briefly exposed to their natural resonant frequency, for example due to rapid changes in a non-sinusoidal field like a square wave field. So it is recommended to use resonant models only for sinusoidal CW fields to avoid distortion of the output due to ringing.  
For screenshots and more measurement info see  www.magneticsciences.com/measurement-notes
All these sensors can also be used for vibration monitoring to measure mechanical vibrations or rotations using the sensor's motion through the earth's magnetic field or through the field of a nearby magnet. 
Each MC series probe is manufactured, tested, and calibrated at our company center in Massachusetts.  Magnetic Sciences is a small business and we define ourselves by quality and integrity.
Brand: Magnetic Sciences
Isin: IBHBMGFLFUMP

$296.19

$296.84
0% off
  • Free shipping in US
  • Arrives:

  • Free 30-Day returns

In Stock

Sold by , Fulfilled by IBSPOT
Ships from USA

IBspot Buyer Protection

Shop confidently on IBspot, receive your item as described or your money back for eligible orders. Learn Program Terms

More seller options

Starting from

Compare all sellers
scroll to top arrow
Magnetic Field Probe model MC110A EMC EMI RFI Calibrated Electromagneti

Magnetic Field Probe model MC110A EMC EMI RFI Calibrated Electromagneti

$296.19 $296.84 (0% off)

Product Overview

The MC110A sensor provides small size and a strong wideband rounded response from 30 kHz to 1 MHz.

Frequency Range:
 
2 kHz - 1 MHz
Shape:
 
B1 Box, small square
Size:
 
1” x 1” x 0.8” (25x25x20mm)
Connector:
 
BNC(f) on box
Weight (approx):
 
1 oz (30g)


About the MC series probes- 
These calibrated magnetic field sensors or probes are cost-effective tools to detect magnetic fields from 5 Hz to 1 MHz (any waveforms), and to accurately measure magnetic field strength of continuous-wave (CW sine wave) fields from 5 Hz to 1 MHz. 
These single-axis "B-dot" coil sensors respond to magnetic fields parallel to the coil axis; they produce an output voltage calibrated to the CW magnetic field strength. You display the output voltage on your own instrument (AC or RF voltmeter, multimeter, oscilloscope, or spectrum analyzer with high input impedance). The frequency of the sensor output voltage is the same as the frequency of the magnetic field. 
NIST traceable calibration certificate with data is printed and shipped with each sensor. No battery or power supply is needed for the sensors. These are easy-to-use affordable sensors for magnetic field measurement, EMC / EMI / RFI testing, and troubleshooting. These sensors do not measure static or DC magnetic fields, for example from motionless magnets or the earth's magnetic field, unless the sensor is vibrating or rotating. 
Frequency Response Graph (see photos): shows each sensor model output voltage (as Volts output per Gauss, or mV output per mG field) at each frequency, when placed in a continuous sine wave (CW) magnetic field at that frequency. Use this graph (or printed calibration data), and the sensor output voltage, to determine the magnetic field in Gauss. The graph is accurate for a continuous sine wave (CW) magnetic field. Individual sensors may vary, so calibration data is printed and shipped with each sensor.
Examples: at your frequency if the Volts per Gauss = 1 (on vertical axis), and you are measuring 80 mV, then the CW field = 80 mG.  (80mV /1 = 80mG)
           Or, at your frequency if the Volts per Gauss = 8 (on vertical axis), and you are measuring 80 mv, then the CW field =10 mG.  (80mV /8  = 10mG)
           Or, at your frequency if the Volts per Gauss = 5 (on vertical axis), and you are measuring 3 V, then the CW field =0.6 G.  (3V /5 = 0.6 G)
If your display instrument is measuring RMS volts then you are measuring RMS magnetic field strength. (RMS is recommended). 
If your display instrument is measuring Peak volts then you are measuring Peak magnetic field strength. 
Unit conversion:  10G = 1mT,  1 mG = 0.1 uT.      1 kA/m  = 1.26 mT  = 12.6 G in air.
To Use the Sensor: Connect the sensor to your display instrument (multimeter, AC or RF voltmeter, spectrum analyzer, or oscilloscope, etc). Place the center of the sensor at the location you want to measure magnetic field strength.
The sensor is single axis and responds to the magnetic field parallel to the sensor axis, which is along the longest dimension of the sensor (parallel to the writing on the sensor label). The sensing area is along most of the length of sensor axis, so if the field is spatially non-uniform it will be averaged over the length of the sensor. 
To find the maximum field polarization, turn the sensor in different directions to see the largest reading, then the sensor axis is parallel to the maximum magnetic field polarization direction. The polarization direction of the field (max reading) is often at right angles to the direction towards the source of the field. 
The reading will also increase as you get closer to the source of the field, although the immediate surroundings, multiple sources, and the sensor orientation (pointing direction) can cause variations. Sometimes you won't see exactly the same reading when you check the same location again, this is usually because the sensor is not exactly at the same location and orientation. Hold the sensor still, because vibrating, shaking, or quick movements can cause extraneous readings due to sensor acceleration through the earth’s static magnetic field.
To measure the magnetic field:  Measure the alternating (AC or RF) voltage output from your sensor, then reference the graph above (or the calibration data delivered with each sensor) at your frequency to determine the magnetic field strength. 
Minimum Measurable Field and Resolution: These are determined by the resolution and noise level of your display instrument. The sensors are passive devices and contribute negligible thermal noise.
Maximum Measurable Field: Sensors may be damaged by magnetic fields producing more than 50 Volts output from the sensor. Also, saturation of the core can cause inaccuracies above 50 Gauss (5 mT) field. The longest sensors (sizes C4 and C5) may start to saturate at 25 Gauss (2.5 mT).  Since these are B-dot sensors, a fast change in the field can cause output voltage to peak which could damage the large sensors MC90R and MC90-110 with high outputs. If unsure then best to gradually ramp-up and down the field strength. 
Temperature Range: Sensors can operate from -30 C to + 55 C (-22 F to +131 F), or in some cases a wider temp range. 
Calibration: Calibration standards and instruments are NIST traceable. Each sensor is individually calibrated using a CW sine wave magnetic field at a number of frequencies, and the calibration data is printed and shipped with each sensor. Calibration is usually accurate for many years for these passive sensors. Calibration results may vary near resonant frequencies and above 500 kHz.  
Instrument Impedance and Coax: To display the calibrated output voltage from the sensor we recommend you use an instrument with high input impedance. The length of coaxial cable you use can also significantly affect readings above 30 kHz, due to coax capacitance, see Impedance  
Sinusoidal and Non-Sinusoidal Fields: These coils are known as "B-dot" sensors since the voltage output is the time-derivative of the magnetic field. So if the magnetic field is a continuous CW sine wave, then the sensor output is a cosine CW voltage at the same frequency, and the response curves in the figure above show the correct output voltage per gauss at each frequency. 
For non-sinusoidal waveforms the output voltage is also the first derivative (B-dot) of the magnetic field. So when exposed to a rapid large change in the magnetic field (like switching the field on or off, or square wave or spike in the field), then the sensor output voltage will show a peak or bump, due to the fast change in the field giving a large first derivative. 
Some sensors are described as "resonant". Their frequency peak is the resonant frequency. Those "resonant" models will "ring" for a few milliseconds if briefly exposed to their natural resonant frequency, for example due to rapid changes in a non-sinusoidal field like a square wave field. So it is recommended to use resonant models only for sinusoidal CW fields to avoid distortion of the output due to ringing.  
For screenshots and more measurement info see  www.magneticsciences.com/measurement-notes
All these sensors can also be used for vibration monitoring to measure mechanical vibrations or rotations using the sensor's motion through the earth's magnetic field or through the field of a nearby magnet. 
Each MC series probe is manufactured, tested, and calibrated at our company center in Massachusetts.  Magnetic Sciences is a small business and we define ourselves by quality and integrity.
Brand: Magnetic Sciences
Isin: IBHBMGFLFUMP

The MC110A sensor provides small size and a strong wideband rounded response from 30 kHz to 1 MHz.

Frequency Range:
 
2 kHz - 1 MHz
Shape:
 
B1 Box, small square
Size:
 
1” x 1” x 0.8” (25x25x20mm)
Connector:
 
BNC(f) on box
Weight (approx):
 
1 oz (30g)


About the MC series probes- 

These calibrated magnetic field sensors or probes are cost-effective tools to detect magnetic fields from 5 Hz to 1 MHz (any waveforms), and to accurately measure magnetic field strength of continuous-wave (CW sine wave) fields from 5 Hz to 1 MHz. 

These single-axis "B-dot" coil sensors respond to magnetic fields parallel to the coil axis; they produce an output voltage calibrated to the CW magnetic field strength. You display the output voltage on your own instrument (AC or RF voltmeter, multimeter, oscilloscope, or spectrum analyzer with high input impedance). The frequency of the sensor output voltage is the same as the frequency of the magnetic field. 

NIST traceable calibration certificate with data is printed and shipped with each sensor. No battery or power supply is needed for the sensors. These are easy-to-use affordable sensors for magnetic field measurement, EMC / EMI / RFI testing, and troubleshooting. These sensors do not measure static or DC magnetic fields, for example from motionless magnets or the earth's magnetic field, unless the sensor is vibrating or rotating. 

Frequency Response Graph (see photos): shows each sensor model output voltage (as Volts output per Gauss, or mV output per mG field) at each frequency, when placed in a continuous sine wave (CW) magnetic field at that frequency. Use this graph (or printed calibration data), and the sensor output voltage, to determine the magnetic field in Gauss. The graph is accurate for a continuous sine wave (CW) magnetic field. Individual sensors may vary, so calibration data is printed and shipped with each sensor.

Examples: at your frequency if the Volts per Gauss = 1 (on vertical axis), and you are measuring 80 mV, then the CW field = 80 mG.  (80mV /1 = 80mG)
           Or, at your frequency if the Volts per Gauss = 8 (on vertical axis), and you are measuring 80 mv, then the CW field =10 mG.  (80mV /8  = 10mG)
           Or, at your frequency if the Volts per Gauss = 5 (on vertical axis), and you are measuring 3 V, then the CW field =0.6 G.  (3V /5 = 0.6 G)
If your display instrument is measuring RMS volts then you are measuring RMS magnetic field strength. (RMS is recommended). 
If your display instrument is measuring Peak volts then you are measuring Peak magnetic field strength. 
Unit conversion:  10G = 1mT,  1 mG = 0.1 uT.      1 kA/m  = 1.26 mT  = 12.6 G in air.

To Use the Sensor: Connect the sensor to your display instrument (multimeter, AC or RF voltmeter, spectrum analyzer, or oscilloscope, etc). Place the center of the sensor at the location you want to measure magnetic field strength.

The sensor is single axis and responds to the magnetic field parallel to the sensor axis, which is along the longest dimension of the sensor (parallel to the writing on the sensor label). The sensing area is along most of the length of sensor axis, so if the field is spatially non-uniform it will be averaged over the length of the sensor. 

To find the maximum field polarization, turn the sensor in different directions to see the largest reading, then the sensor axis is parallel to the maximum magnetic field polarization direction. The polarization direction of the field (max reading) is often at right angles to the direction towards the source of the field. 

The reading will also increase as you get closer to the source of the field, although the immediate surroundings, multiple sources, and the sensor orientation (pointing direction) can cause variations. Sometimes you won't see exactly the same reading when you check the same location again, this is usually because the sensor is not exactly at the same location and orientation. Hold the sensor still, because vibrating, shaking, or quick movements can cause extraneous readings due to sensor acceleration through the earth’s static magnetic field.

To measure the magnetic field:  Measure the alternating (AC or RF) voltage output from your sensor, then reference the graph above (or the calibration data delivered with each sensor) at your frequency to determine the magnetic field strength. 

Minimum Measurable Field and Resolution: These are determined by the resolution and noise level of your display instrument. The sensors are passive devices and contribute negligible thermal noise.

Maximum Measurable Field: Sensors may be damaged by magnetic fields producing more than 50 Volts output from the sensor. Also, saturation of the core can cause inaccuracies above 50 Gauss (5 mT) field. The longest sensors (sizes C4 and C5) may start to saturate at 25 Gauss (2.5 mT).  Since these are B-dot sensors, a fast change in the field can cause output voltage to peak which could damage the large sensors MC90R and MC90-110 with high outputs. If unsure then best to gradually ramp-up and down the field strength. 

Temperature Range: Sensors can operate from -30 C to + 55 C (-22 F to +131 F), or in some cases a wider temp range. 

Calibration: Calibration standards and instruments are NIST traceable. Each sensor is individually calibrated using a CW sine wave magnetic field at a number of frequencies, and the calibration data is printed and shipped with each sensor. Calibration is usually accurate for many years for these passive sensors. Calibration results may vary near resonant frequencies and above 500 kHz.  

Instrument Impedance and Coax: To display the calibrated output voltage from the sensor we recommend you use an instrument with high input impedance. The length of coaxial cable you use can also significantly affect readings above 30 kHz, due to coax capacitance, see Impedance  

Sinusoidal and Non-Sinusoidal Fields: These coils are known as "B-dot" sensors since the voltage output is the time-derivative of the magnetic field. So if the magnetic field is a continuous CW sine wave, then the sensor output is a cosine CW voltage at the same frequency, and the response curves in the figure above show the correct output voltage per gauss at each frequency. 

For non-sinusoidal waveforms the output voltage is also the first derivative (B-dot) of the magnetic field. So when exposed to a rapid large change in the magnetic field (like switching the field on or off, or square wave or spike in the field), then the sensor output voltage will show a peak or bump, due to the fast change in the field giving a large first derivative. 

Some sensors are described as "resonant". Their frequency peak is the resonant frequency. Those "resonant" models will "ring" for a few milliseconds if briefly exposed to their natural resonant frequency, for example due to rapid changes in a non-sinusoidal field like a square wave field. So it is recommended to use resonant models only for sinusoidal CW fields to avoid distortion of the output due to ringing.  

For screenshots and more measurement info see  www.magneticsciences.com/measurement-notes

All these sensors can also be used for vibration monitoring to measure mechanical vibrations or rotations using the sensor's motion through the earth's magnetic field or through the field of a nearby magnet. 

Each MC series probe is manufactured, tested, and calibrated at our company center in Massachusetts.  Magnetic Sciences is a small business and we define ourselves by quality and integrity.
  • Brand: Magnetic Sciences
  • MPN: MC110A
  • Country/Region of Manufacture: United States

 

Processing Time

  • We process and ship orders Monday through Friday, excluding holidays.
  • Most orders are processed within 1–3 business days, depending on product availability.
  • Processing time does not include delivery time, which varies based on your location and the product’s shipping origin.

Shipping Coverage

We currently ship to addresses within the United States only.

Products may ship from:

  • Our U.S. warehouses, or
  • Our international fulfillment centers (including Korea, Japan, China, Australia, Mexico, Canada, the United Kingdom, and the EU).

Each product page clearly displays the “Ships From” location before checkout.


Shipping Methods & Estimated Delivery

 

Shipping Method Estimated Delivery Cost
Free Shipping (US) 5–10 business days Free, available on eligible items
Standard Shipping (US to US by Item) 5–10 business days $5.95 for the first item, $3.00 each additional
Standard Shipping (From US Warehouse by Weight) 3–7 business days Starts at $5.95, increasing by $2 for every additional 4 oz (0.25 lb) up to 1 lb, then by $4 for each additional pound beyond 1 lb.
Extended Delivery (Ships from Overseas Warehouse) 10–20 business days $14.50 for the first item, $2.00 each additional

Shipping costs vary by product weight, quantity and origin. Exact rates and delivery estimates are displayed on each product page and confirmed during checkout.

Free Shipping

We offer free standard shipping on select products or during promotional periods. Availability of free shipping will be displayed on the product page and at checkout. 


Carriers

We primarily use USPS and UPS for U.S. deliveries. Items shipped from international warehouses may be handled by partner carriers depending on the country of origin.


Import Duties & Taxes

All import duties and taxes for goods imported into the United States are included in the product price. There are no hidden fees or additional charges at checkout.


Tracking & Notifications

Every order is fully trackable. You will receive a tracking number via email once your order ships. If your order ships in multiple packages, each package will have its own tracking number. If you don’t receive a shipping confirmation immediately, your order is still being processed and will arrive within the estimated timeframe shown at checkout.

You can track your shipment anytime using the Track Your Order page on our website.


Order Changes & Cancellations

If you need to modify or cancel an order, please contact our customer support team as soon as possible. Orders can only be canceled before they ship. Once shipped, cancellations are not possible, but you may request a return after the item is delivered.

Contact: support@ibspot.com


Item Not Received

If your tracking number shows “Delivered” but you haven’t received your package:

  1. Check with your local USPS or UPS office.
  2. Confirm that the shipping address provided was correct.
  3. If the issue remains unresolved, contact support@ibspot.com for assistance.

Damaged or Lost Parcels

If your package arrives damaged or fails to arrive, please contact us immediately. For deliveries to P.O. boxes, ibspot is not responsible for damage caused by weather, temperature, or theft.


Customer Support

For any questions about shipping, tracking, or delivery, our customer support team is here to help.

Email: support@ibspot.com

We’re always happy to help.

Please read our policy carefully before making a purchase.
We aim to ensure every customer has a smooth and transparent experience with ibspot.com.


Order Cancellation Policy

Customers may request to cancel an order before it has been shipped.
Once the package has been shipped, cancellations are no longer possible; however, you may still request a return after receiving your order.

How to Request a Cancellation

You can contact us through:

Please submit your cancellation request as soon as possible after placing the order to allow us to process it before shipment.


Return Policy

We accept returns for most items within 30 days of delivery for a full refund.

To be eligible for a return:

  • The item must be unused, in the same condition as received, and in its original packaging.
  • A tracking number must be provided to confirm the return shipment.

Non-Returnable Items

Certain products cannot be returned, including:

  • Perishable goods (for example, food, flowers, newspapers, magazines)
  • Intimate or sanitary goods
  • Hazardous materials or flammable liquids/gases
  • Gift cards
  • Downloadable software
  • Some health and personal care items

Partial Refunds (If Applicable)

Partial refunds may be granted in specific situations, such as:

  • Books with obvious signs of use
  • Opened CDs, DVDs, software, or vinyl records
  • Items not in their original condition, damaged, or missing parts not due to our error
  • Items returned more than 30 days after delivery

How to Return an Item

To initiate a return, please contact us at support@ibspot.com with your order number and details about the product you wish to return.
Our team will provide you with return instructions and a prepaid return label.


Shipping Cost for Returns

Please contact us before returning any item.
We will provide a free return shipping label.

If a return is sent back without prior contact or without our provided label, we cannot be held responsible for return shipping costs.

For items valued over $75, we recommend using a trackable shipping service or purchasing shipping insurance.
We cannot guarantee that we will receive your returned item if shipped independently.


Refund Processing

Once your return is received and inspected:

  • In-store returns are refunded to the original form of payment or issued as a gift card.
  • Mail-in returns using our prepaid label are refunded to the original payment method within 3–5 business days after we receive your return.

Please allow:

  • 5–7 business days for the return to reach our Returns Center.
  • An additional 3–10 business days for your bank to post the refund to your account.

Damages, Issues & Claims

Please inspect your order immediately upon receipt.
If you receive a defective, damaged, or incorrect item, contact us right away at support@ibspot.com.
We’ll evaluate the issue promptly and make it right.

Product Claims

Before purchasing, please review product details carefully.
If there is a problem with your order upon arrival, visit our Support Center or contact us directly to arrange return shipping or replacement.


Exchanges

We do not process direct exchanges.
The fastest way to get what you need is to return the original item and place a new order once your return is accepted.


Return Address

IBSPOT Return Center
15 Sawmill Ln
Dover Plains, NY 12522
United States


Customer Support

Our support team is available 24/7 to assist with cancellations, returns, or general inquiries.

Email: support@ibspot.com
Contact Form: Contact Us page

Oops!

Sorry, it looks like some products are not available in selected quantity.

OK

Subscribe To Our Newsletter

Disclaimer: Statements made, or products sold through this website, have not been evaluated by the United States Food and Drug Administration. Products are not intended to diagnose, treat, cure or prevent any disease.